Surgical Correction of Spontaneous Outflow Graft Twisting of HeartMate III via a Subc...
Hidefumi Nishida
Valluvan Jeevanandam

Hidefumi Nishida

and 2 more

August 05, 2020
We experienced two cases of successful surgical correction of HeartMate 3 outflow graft twisting through a subcostal approach. They were diagnosed with computed tomography or pull back pressure measurement. Technically, a subcostal approach allowed us to access directly to the twisted outflow graft and the device connector in a less invasive fashion without a re-sternotomy. Diagnostic modality and surgical tips are presented.
Successful endovascular repair of aortic rupture caused by axillary intra-aortic ball...
Hidefumi Nishida
Tae Song

Hidefumi Nishida

and 4 more

May 20, 2020
We report the successful endovascular repair of a rare case of aortic rupture caused by axillary intra-aortic balloon pump insertion failure. A 38-year-old Jehovah’s Witness female with situs inversus totalis was referred to our hospital for acute decompensated heart failure. We placed an axillary intra-aortic balloon pump for circulatory support. However, an exchange was required due to balloon malfunction (kinked driveline). Unfortunately, the exchange was complicated by an iatrogenic aortic rupture along with large hematoma compressing the trachea. Emergent endovascular repair was performed successfully without any blood transfusion. Postoperative computed tomography showed a successfully repaired aorta and resolving hematoma.