Retrieval of a Leadless Transcatheter Pacemaker from the Tricuspid Valve Annulus: A Case Report
  • Jiahui Chen,
  • Xueying Chen,
  • YANGANG SU
Jiahui Chen
Zhongshan Hospital Fudan University

Corresponding Author:chen.jiahui1@zs-hospital.sh.cn

Author Profile
Xueying Chen
Zhongshan Hospital Fudan University
Author Profile
YANGANG SU
Zhongshan Hospital Fudan University
Author Profile

Abstract

This case report describes a procedure of retrieval of a leadless transcatheter pacemaker from the right ventricle after device implantation immediately. An 80-year-old man affected by ischemic cardiomyopathy, complete AV block and atrial fibrillation was implanted with a Micra transcatheter pacing system at the median septum of the right ventricle. After tether removal, the leadless pacemaker migrated to tricuspid valve annulus. The device was successfully removed using a snare loop hooked to the proximal retrieval feature of Micra.