Left Ventricular Assist Device Implantation via Bi-Thoracotomy
Technique: A Single-Center Perspective
Abstract
Background: Left ventricular assist devices (LVAD) are standardly
implanted via full sternotomy. Non-sternotomy approaches are gaining
popularity, but potential benefits of this approach have not been
well-studied. We hypothesized that LVAD implantation by bi-thoracotomy
(BT) would demonstrate smaller and more consistent inflow cannula angles
leading to improved postoperative outcomes compared to sternotomy.
Methods: Charts of patients who underwent LVAD implantation between June
2018 and June 2020 at a single academic institution were retrospectively
reviewed. Patient demographics, surgical approach (sternotomy vs. BT),
laboratory values, and postoperative course were compared. The inflow
cannula angle was measured on the first chest radiograph available
postoperatively. Results: Of 40 patients studied, BT approach was used
in 17 (42.5%). Mean inflow cannula angles were smaller in BT patients
(23.0 vs. 37.1 degrees, p=0.018) and had a smaller standard deviation
(13.8 vs. 20.3). Excluding patients who went on to receive heart
transplant or died in the same hospitalization, there was no difference
in median length of hospital stay after surgery (16.0 vs. 17.5 days,
p=0.768). However, BT patients required fewer days of postoperative
inotrope support (4.0 vs. 7.0 days, p=0.012). Conclusions: Our data
suggest inflow cannula angles are smaller and more consistent with the
BT approach, which leads to shorter duration of postoperative inotropic
support. This finding may suggest improved right heart function
following LVAD implant via BT approach. Further study is warranted to
determine additional benefits of the BT approach.