loading page

Repositioning and Extraction of Stylet-Driven Pacing Leads with Extendable Helix used for Left Bundle Branch Area Pacing.
  • +7
  • Jean-Benoit le Polain de Waroux,
  • Jean-Yves Wielandts,
  • Kris Gillis,
  • Gabriela Hilfiker,
  • Antonio Sorgente,
  • Lucio Capulzini,
  • Bert Geerts,
  • Sebastien Knecht,
  • Mattias Duytschaever,
  • Rene Tavernier
Jean-Benoit le Polain de Waroux
AZ Sint-Jan Brugge-Oostende AV

Corresponding Author:jblpdw@hotmail.com

Author Profile
Jean-Yves Wielandts
AZ Sint-Jan Brugge-Oostende AV
Author Profile
Kris Gillis
AZ Sint-Jan Brugge-Oostende AV
Author Profile
Gabriela Hilfiker
AZ Sint-Jan Brugge-Oostende AV
Author Profile
Antonio Sorgente
Hospital Centre EpiCURA Hornu
Author Profile
Lucio Capulzini
Hospital Centre EpiCURA Hornu
Author Profile
Bert Geerts
AZ Sint-Jan Brugge-Oostende AV
Author Profile
Sebastien Knecht
AZ Sint-Jan Brugge-Oostende AV
Author Profile
Mattias Duytschaever
AZ Sint-Jan Brugge-Oostende AV
Author Profile
Rene Tavernier
AZ Sint-Jan Brugge-Oostende AV
Author Profile

Abstract

Abstract: Conventional stylet-driven leads with extendable helix can be implanted successfully for Left Bundle Branch Area Pacing (LBBAP) with a low acute complication rate. However, removal and repositioning techniques of these leads haven’t been yet described. We report 2 cases in which lead repositioning after a first unsuccessful attempt to LBBAP was associated with fracture of the helix rotating mechanism and failure to fully extract the pacing lead.
10 Jan 2021Submitted to Journal of Cardiovascular Electrophysiology
13 Jan 2021Submission Checks Completed
13 Jan 2021Assigned to Editor
24 Jan 2021Reviewer(s) Assigned
09 Feb 2021Review(s) Completed, Editorial Evaluation Pending
15 Feb 2021Editorial Decision: Revise Minor
27 Feb 20211st Revision Received
27 Feb 2021Submission Checks Completed
27 Feb 2021Assigned to Editor
27 Feb 2021Reviewer(s) Assigned
04 Mar 2021Review(s) Completed, Editorial Evaluation Pending
08 Mar 2021Editorial Decision: Accept