Optimization of superior vena cava isolation with aid of ablation index guidance
  • +3
  • Jun Liu,
  • JInrui Guo,
  • Xiaofeng Li,
  • Yu Xia,
  • Guo-Dong Niu,
  • Yan Yao
Jun Liu
Fuwai Hospital State Key Laboratory of Cardiovascular Disease

Corresponding Author:liujundoctor@163.com

Author Profile
JInrui Guo
Fuwai Yunnan Cardiovascular Hospital
Author Profile
Xiaofeng Li
Fuwai Hospital State Key Laboratory of Cardiovascular Disease
Author Profile
Yu Xia
Fuwai Hospital State Key Laboratory of Cardiovascular Disease
Author Profile
Guo-Dong Niu
Fuwai Yunnan Cardiovascular Hospital
Author Profile
Yan Yao
Fuwai Hospital State Key Laboratory of Cardiovascular Disease
Author Profile

Abstract

Objective: To investigate the optimal range of quantitative ablation index (AI) value during superior vena cava (SVC) electrical isolation by radiofrequency catheter ablation (RFCA). Methods: First, in a development cohort of patients with atrial fibrillation (AF), the RFCA with 40W was performed to complete SVC isolation guided by the conduction breakthrough point from the right atrium to SVC. Then, the range of AI value was calculated by offline analysis on different segments of SVC. Lastly, for the validation of AF patients, the safety and effectiveness of SVC isolation with the optimized target range of AI value were evaluated with an additional adenosine test. Results: A total of 101 patients with AF were included in the study (44 patients in the development cohort / 57 in the validation cohort). The segmental ablation strategy was applied in 70% of the patients. According to the offline analysis of the AI values in the development cohort, the target AI value range was set as 350-400. The success rate of SVC isolation in the validation cohort was significantly higher than that in the exploration cohort (100% vs 90.9%, P = 0.02), and no complications occurred in the exploration cohort. During the adenosine test, the recovery rate of electrical conduction in SVC was significantly lower than that in the pulmonary vein (3.5% vs 17.5%). Conclusion: The target AI value with a range from 350 to 400 is safe and effective for high-power RFCA to complete SVC isolation.
16 Apr 2023Submitted to Journal of Cardiovascular Electrophysiology
16 Apr 2023Submission Checks Completed
16 Apr 2023Assigned to Editor
16 Apr 2023Review(s) Completed, Editorial Evaluation Pending
19 Apr 2023Reviewer(s) Assigned
25 May 2023Editorial Decision: Revise Minor
23 Jun 20231st Revision Received
30 Jun 2023Review(s) Completed, Editorial Evaluation Pending
30 Jun 2023Submission Checks Completed
30 Jun 2023Assigned to Editor
30 Jun 2023Reviewer(s) Assigned
06 Jul 2023Editorial Decision: Accept